Cooling rate | 4 °C/min | 4 °C/min | 4 °C/min | 4 °C/min |
Gripper carrying capacity | 1,200 g | 1,200 g | 1,200 g | 1,200 g |
Heating rate | 9 °C/min | 9 °C/min | 9 °C/min | 9 °C/min |
Operation | Electronic | Electronic | Electronic | Electronic |
Pipetting type | Air-cushion | Air-cushion | Air-cushion | Air-cushion |
Noise level | ~ 56 dB(A) | ~ 56 dB(A) HEPA off | ~ 56 dB(A) | ~ 56 dB(A) |
Application | PCR/qPCR and normalization | PCR/qPCR and normalization | PCR/qPCR and normalization | PCR/qPCR and normalization |
Volume range | 0.2 µL – 1 mL | 0.2 µL – 1 mL | 0.2 µL – 1 mL | 0.2 µL – 1 mL |
Power supply | 100 – 240 V ±10 %, 50 – 60 Hz ±5 % | 100 – 240 V ±10 %, 50 – 60 Hz ±5 % | 100 – 240 V ±10 %, 50 – 60 Hz ±5 % | 100 – 240 V ±10 %, 50 – 60 Hz ±5 % |
Max. power consumption | 600 W | 600 W | 600 W | 600 W |
Dimensions (W × D × H) | 65 × 61 × 67 cm / 26 × 24 × 27 in | 65 × 61 × 77 cm / 26 × 24 × 30 in | 65 × 61 × 67 cm / 26 × 24 × 27 in | 65 × 61 × 77 cm / 26 × 24 × 30 in |
Weight w/o accessories | 50 kg / 110.2 lb | 55 kg / 121.3 lb | 50 kg / 110.2 lb | 55 kg / 121.3 lb |
X-Y-Z axis positioning accuracy | | | | |
Random deviation | ±0.3 mm | ±0.3 mm | ±0.3 mm | ±0.3 mm |
Systematic deviation | ±0.1 mm | ±0.1 mm | ±0.1 mm | ±0.1 mm |
Eppendorf EasyCon | | | | |
Display | LCD color display | LCD color display | LCD color display | LCD color display |
Display resolution | 1,024 × 600 pixels | 1,024 × 600 pixels | 1,024 × 600 pixels | 1,024 × 600 pixels |
Interfaces | 4 × USB 2.0 | 4 × USB 2.0 | 4 × USB 2.0 | 4 × USB 2.0 |
Touch panel | Capacitive touch panel | Capacitive touch panel | Capacitive touch panel | Capacitive touch panel |
Thermal module (optional) | | | | |
Temperature range | 0 °C – 110 °C | 0 °C – 110 °C | 0 °C - 110 °C | 0 °C - 110 °C |
Conductor – X, Y, Z Positioning | | | | |
Random measurement error | ±0.1 mm | ±0.1 mm | ±0.1 mm | ±0.1 mm |
Detector | | | | |
Optical confocal infrared detector | Contact-free recognition of fill level, inserted tools, labware, tip types and quantities | Contact-free recognition of fill level, inserted tools, labware, tip types and quantities | Contact-free recognition of fill level, inserted tools, labware, tip types and quantities | Contact-free recognition of fill level, inserted tools, labware, tip types and quantities |
UV lamp | | | | |
Irradiation period per run | – | 15 min | – | – |
Main wavelength | – | 254 nm (UV-C) | – | – |
Radiation capacity after 100 h | – | 3,9 W | – | – |
Radiation capacity at a distance of 1 m | – | 39 µW/cm2 | – | – |
HEPA filter | | | | |
Filter class according to EN 1822 | – | E 10 | – | – |
Volume flow | – | 70 – 80 m3/h | – | – |
Dispensing tool | | | | |
Random measurement error (1 µL)1,2 | ≤3 % | ≤3 % | ≤3 % | ≤3 % |
Random measurement error (1,000 µL)1,2 | ≤0.15 % | ≤0.15 % | ≤0.15 % | ≤0.15 % |
Random measurement error (50 µL)1,2 | ≤0.4 % | ≤0.4 % | ≤0.4 % | ≤0.4 % |
Systematic measurement error (1 µL)1,2 | ±5 % | ±5 % | ±5 % | ±5 % |
Systematic measurement error (1,000 µL)1,2 | ±0.7 % | ±0.7 % | ±0.7 % | ±0.7 % |
Systematic measurement error (50 µL)1,2 | ±1.2 % | ±1.2 % | ±1.2 % | ±1.2 % |